6
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
Device care
Clean the device only with a soft, dry cloth.
Accuracy test
We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). The verification time
period is 2 years. Verification is carried out in accordance with the
document Р 1323565.2.001-2018 «State system for ensuring the
uniformity of measurements. Non-invasive blood pressure meters.
Verification Method» Confirmation of initial verification, an elec-
tronic version of the verification certificate, is on the website of the
Federal Information Fund for Ensuring Uniformity of Measure-
ments - www.fgis.gost.ru
Information on verification is on the company website -
www.microlife.ru
Disposal
When the disposal of this device is required please contact a
specialized organization, authorized to provide utilization
according to the legislation of the Russian Federation.
11.Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
Batteries, cuff and parts that become worn with use are not
included.
Opening or altering the device invalidates the guarantee.
The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
Please contact your local Microlife-Service (see foreword).
12. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Operating conditions: 10 - 40 °C / 50 - 104 °F
30 - 85 % relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
30 - 85 % relative maximum humidity
Weight: 400 g (including batteries)
Dimensions: 140 x 120 x 70 mm
Measuring procedure: oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V
diastolic
Measurement range: 20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range: 0 - 299 mmHg
Resolution: 1 mmHg
Static accuracy: pressure within ± 3 mmHg
Pulse accuracy: ± 5 % of the readout value
Voltage source:
4 x 1.5 V alkaline batteries; size AA
Mains adapter DC 6V, 600 mA (optional)
Battery lifetime: approx. 920 measurements
(using new batteries)
Included accessories:
Cuff
Tube
Instruction manual
Warranty card
Batteries
Power Adapter
Packaging Box
IP Class: IP20
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years

Table of Contents

Other manuals for Microlife BP 3AG1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP 3AG1 and is the answer not in the manual?

Microlife BP 3AG1 Specifications

General IconGeneral
TypeAutomatic
Measuring methodOscillometric
Speaking functionNo
Placement supportedUpper arm
Mean arterial pressureYes
Battery typeAA
Battery voltage1.5 V
Number of batteries supported4
Display typeLCD

Related product manuals