9. Technical specifications:
Weight: 265 g (with batteries
Size: 115 (W) x 181 (L) x 76 (H) mm
Storage temperature: -20 - +55 °C / -4 - +131 °F; 15 - 95 % relative maximum humidity
Operation temperature: 10 - 40 °C / 50 - 104 °F; 15 - 95 % relative maximum humidity
Display: LCD-Display (Liquid Crystal Display)
Measuring method: oscillometric
Pressure sensor: capacitive
Measuring range:
SYS/DIA: 30 to 280 mmHg
Pulse: 40 to 200 per minute
Cuff pressure display range: 0–299 mmHg
Memory: Storing the last measurements automatically
Measuring resolution: 1 mmHg
Accuracy:
Pressure within ± 3 mmHg
Pulse ± 5 % of the reading
Power source: 4 dry cells (batteries) UM-3, size AA, 1.5 V
Cuffs: type M for arm circumference 22–32 cm or
type L for arm circumference 32–42 cm (optional)
Reference to standards: EN 1060-1 /-3 /-4; IEC 60601-1; IEC 60601-1-2 (EMC)
This device complies with the requirements of the Medical Device Directive 93/42/EEC.
Technical alterations reserved.
10. www.microlife.com
Detailed user information about our thermometers and blood pressure monitors as well as services can be
found at www.microlife.com.