8. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
* Please consult your doctor, if this or any other problem occurs repeatedly.
9. Safety, Care, Accuracy Test and Disposal
Safety and protection
Instrument care
Clean the instrument only with a soft, dry cloth.
Cleaning the cuff
You can machine wash the cuff cover at 30°C (do not iron!).
Accuracy test
We recommend this instrument is tested for accuracy every
2 years or after mechanical impact (e.g. being dropped). Please
contact Microlife-Service to arrange the test (see foreword).
Disposal
Error Description Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat the
measurement.*
«
ERR 2
» Error signal
During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be generated
in the cuff. A leak may have occurred.
Check that cuff and pump ball are correctly
connected and that the cuff is not fitted too
loosely. Replace the batteries if necessary.
Repeat the measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and
no result can therefore be displayed. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
«
HI
» Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
«
LO
» Pulse too low The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*
)
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
This instrument may be used only for the purpose described in
this booklet. The manufacturer cannot be held liable for damage
caused by incorrect application.
This instrument comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section!
Protect it from:
water and moisture
extreme temperatures
impact and dropping
contamination and dust
direct sunlight
heat and cold
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use the instrument close to strong electromagnetic fields
such as mobile telephones or radio installations.
Do not use the instrument if you think it is damaged or notice
anything unusual.
Never open the instrument.
If the instrument is not going to be used for a prolonged period
the batteries should be removed.
Read the additional safety instructions in the individual sections
of this booklet.
Ensure that children do not use the instrument unsuper-
vised; some parts are small enough to be swallowed.
WARNING: Under no circumstances, however, may you
wash the inner bladder! Always remove the sensitive
bladder from the sleeve before washing and replace it care-
fully again afterwards.
Batteries and electronic instruments must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.

Other manuals for Microlife BP A80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP A80 and is the answer not in the manual?

Microlife BP A80 Specifications

General IconGeneral
BrandMicrolife
ModelBP A80
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals