Using rechargeable batteries
You can also operate this instrument using rechargeable batteries.
8. Error Messages
If an error occurs during the measurement, the measurement is inter-
rupted and an error message, e.g. «ERR 3», is displayed.
)
The memory retains all values although date and time (and
possibly also set alarm times) must be reset – the year number
therefore flashes automatically after the batteries are replaced.
)
Use 2 new, long-life 1.5V, size AAA batteries.
)
Do not use batteries beyond their date of expiry.
)
Remove batteries if the instrument is not going to be used for
a prolonged period.
)
Only use «NiMH» type reusable batteries.
)
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
instrument as they may become damaged (total discharge as
a result of low use of the instrument, even when switched off).
)
Always remove the rechargeable batteries if you do not intend
to use the instrument for a week or more.
)
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
Error Description Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too weak.
Re-position the cuff and repeat the
measurement.*
«
ERR 2
» Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be generated
in the cuff. A leak may have occurred. Check
that the cuff is correctly connected and is
not too loose. Replace the batteries if neces-
sary. Repeat the measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and no
result can therefore be displayed. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*