IB AG1-20 N-V6 0813
Europe / Middle-East / Africa
Microlife AG
Espenstrasse 139
9443 Widnau / Switzerland
Tel. +41 / 71 727 70 30
Fax +41 / 71 727 70 39
Email admin@microlife.ch
www.microlife.com
Asia
Microlife Corporation.
9F, 431, RuiGang Road, NeiHu
Taipei, 11492, Taiwan, R.O.C.
Tel. 886 2 8797-1288
Fax.886 2 8797-1283
Email service@microlife.com.tw
www.microlife.com
North / Central / South America
Microlife USA, Inc.
1617 Gulf to Bay Blvd., 2nd Floor Ste A
Clearwater, FL 33755 / USA
Tel. +1 727 442 5353
Fax +1 727 442 5377
Email msa@microlifeusa.com
www.microlife.com
Read the instructions carefully before using this device.
Prieš naudodamiesi prietaisu perskaitykite instrukciją.
Pirms šīs ierīces izmantošanas uzmanīgi izlasiet instrukcijas.
Enne seadme kasutust lugege hoolikalt juhiseid.
Перед использованием прибора внимательно прочтите данное руководство.
Lue ohjeet huolellisesti ennen kuin käytät laitetta.

Table of Contents

Other manuals for Microlife BP AG1-20

Questions and Answers

  • A
    Austin CamachoAug 13, 2025
    What to do if the pressure does not rise while pumping Microlife Blood Pressure Monitor?
    • K
      Katherine LopezAug 13, 2025
      If the pressure isn't rising when you pump the bulb on your Microlife Blood Pressure Monitor, ensure the valve is closed. Also, check that the cuff is properly connected to both the bulb and the manometer. Inspect the cuff, tube, and bulb for any leaks. If you find any defective parts, replace them.
  • J
    Julie ReedAug 17, 2025
    How to fix poor sound transmission on Microlife BP AG1-20 Blood Pressure Monitor?
    • J
      jason43Aug 17, 2025
      If you're experiencing poor or distorted sound transmission, or extraneous noise with your Microlife Blood Pressure Monitor, first check the earpieces for dirt or cracks and ensure they are worn correctly. Next, inspect the tube for breaks or twists, and examine the chest piece for damage. Make sure the chest piece is in contact with the skin and positioned over the brachial artery. Clean or replace any defective parts to ensure accurate readings.
  • A
    aprilmoonAug 20, 2025
    What to do if the pressure doesn't rise when pumping the bulb on my Microlife BP AG1-20?
    • C
      Crystal GarnerAug 20, 2025
      If the pressure isn't rising when you pump the bulb, ensure the valve is closed. Also, check that the cuff is properly connected to both the bulb and the manometer. Inspect the cuff, tube, and bulb for any leaks. If you find any defective parts, replace them.
  • A
    Ariana WallerAug 25, 2025
    Why can't I set the deflation rate on my Microlife BP AG1-20?
    • K
      Kristina VaughnAug 25, 2025
      If you cannot set the deflation rate to 2-3 mmHg/sec by adjusting the valve on your Microlife Blood Pressure Monitor, disassemble the valve from the pump ball and check for any blockages in the valve's airway. Clean any blockages and try again. If the issue persists, replace the valve to avoid inaccurate readings.
  • J
    Jessica MillerAug 27, 2025
    How to fix deflation rate if it cannot be set to 2-3 mmHg/sec on Microlife BP AG1-20 Blood Pressure Monitor?
    • L
      Laura BrownAug 28, 2025
      If the deflation rate cannot be set to 2-3 mmHg/sec, disassemble the valve from the bulb and check for any blockages in the airway of the valve. Clean any blockages and try again. If the issue persists, replace the valve to prevent inaccurate readings.
  • G
    gkingAug 31, 2025
    What to do if pointer is not at 0 +/- 3 mmHg at rest on Microlife Blood Pressure Monitor?
    • S
      Stacy WallaceAug 31, 2025
      Ensure that the valve is open when checking if the pointer is at zero. If the deviation is still more than 3 mmHg, contact your dealer to recalibrate the manometer.
  • B
    Brianna MasonSep 4, 2025
    What to do if manometer needle is not at 0 on Microlife BP AG1-20 Blood Pressure Monitor?
    • A
      arielbellSep 4, 2025
      If the manometer needle on your Microlife Blood Pressure Monitor is not at 0 ± 3 mmHg at rest, ensure that the valve is completely open for the zero check. If the deviation is still more than 3 mmHg, contact your dealer to recalibrate the manometer.
  • D
    dowensSep 8, 2025
    Why is the sound transmission poor on my Microlife BP AG1-20 Blood Pressure Monitor?
    • P
      Paul Robinson DDSSep 8, 2025
      Poor sound transmission, distortion, or extraneous noise may be due to several factors. First, check the ear pieces to ensure they are properly plugged in and not cracked or worn. Make sure they fit well. Next, inspect the tube for breaks or twists. Examine the bell and diaphragm of the chest piece for cracks. Ensure the chest piece makes proper contact with the skin and is positioned over the brachial artery during measurement. Clean or replace any defective parts to avoid inaccurate readings.

Summary

Introduction

Features

Details the characteristics and benefits of the aneroid blood pressure kit, including its reliability and portability.

Important Information about Self-Measurement

Emphasizes that self-measurement is for control, not diagnosis or treatment, and advises consulting a doctor for unusual values.

Blood Pressure Information

Causes of High;Low Blood Pressure

Explains the physiological regulation of blood pressure by the brain and circulatory system, including systolic and diastolic values.

Normal Blood Pressure Values

Classifies blood pressure ranges (too low, optimum, normal, slightly high, too high, dangerously high) with recommendations.

Managing Increased;Low Blood Pressure

Provides lifestyle advice and considerations for managing blood pressure, including diet, habits, and physical activity.

Aneroid Blood Pressure Kit Components

Performing a Blood Pressure Measurement

Measurement Preparation

Outlines essential preparation steps, including avoiding exertion, eating, and smoking before taking a measurement.

Common Measurement Errors

Identifies frequent mistakes that can lead to inaccurate blood pressure readings, such as improper cuff fit or patient position.

Correct Cuff Fitting

Provides detailed instructions on correctly positioning and tightening the blood pressure cuff on the arm for accurate readings.

Measurement Procedure Steps

Details the step-by-step process for inflating the cuff, listening for Korotkoff sounds, and recording systolic and diastolic values.

Troubleshooting and Device Malfunctions

Device Care, Maintenance, and Recalibration

Product Guarantee Information

Device Standards and Compliance

Technical Specifications and Data

Microlife BP AG1-20 Specifications

General IconGeneral
TypeUpper Arm Blood Pressure Monitor
Measurement MethodOscillometric
DisplayLCD
Irregular Heartbeat DetectionYes
WHO IndicatorYes
Date and Time DisplayYes
Cuff Size22-32 cm
Memory30 measurements
Power Source4 x AA batteries
User InterfaceButtons
Pulse Rate Range40 - 200 bpm
Intended UseNon-invasive measurement of blood pressure and pulse rate in adults