EasyManua.ls Logo

Microlife BP AG1-20 - Device Standards and Compliance; Technical Specifications and Data

Microlife BP AG1-20
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
8. Reference to standards
Device standard: Device corresponds to the requirements of the
EN 1060-1/-2
ANSI / AAMI SP09
This device complies with the requirements of the Medical Device Directive 93/42/EEC.
9. www.microlife.com
Detailed user information about our products as well as services can be found at www.microlife.com
10. Technical specifications
Weight: 450 g
Size: 175 x 70 x 103 mm
Storage temperature: -20 °C to +70 °C; 85 % relative humidity maximum
Operation temperature: 0 - 46 °C
Measuring range: 0 - 300 mmHg
Measuring resolution: 2 mmHg
Accuracy: within ±3 mmHg in 18 - 33 °C;
within ±6 mmHg in 34 - 46 °C
Inflation source: a volume of at least 200cc to a pressure of 300 mmHg in 4 - 10 sec.
Pressure reduction rate: 2 - 3 mmHg/sec.
Air leakage: < ± 4 mmHg/min
Hysteresis error: within 0 - 4 mmHg
Accessories: 1. M-cuff (adult size with arm circumference of 22 - 32 cm) with
inlaid bladder
2. bulb and valve
3. stethoscope
4. soft bag
Technical alterations reserved.