11
8. Reference to standards
Device standard: Device corresponds to the requirements of the
EN 1060-1/-2
ANSI / AAMI SP09
This device complies with the requirements of the Medical Device Directive 93/42/EEC.
9. www.microlife.com
Detailed user information about our products as well as services can be found at www.microlife.com
10. Technical specifications
Weight: 450 g
Size: 175 x 70 x 103 mm
Storage temperature: -20 °C to +70 °C; 85 % relative humidity maximum
Operation temperature: 0 - 46 °C
Measuring range: 0 - 300 mmHg
Measuring resolution: 2 mmHg
Accuracy: within ±3 mmHg in 18 - 33 °C;
within ±6 mmHg in 34 - 46 °C
Inflation source: a volume of at least 200cc to a pressure of 300 mmHg in 4 - 10 sec.
Pressure reduction rate: 2 - 3 mmHg/sec.
Air leakage: < ± 4 mmHg/min
Hysteresis error: within 0 - 4 mmHg
Accessories: 1. M-cuff (adult size with arm circumference of 22 - 32 cm) with
inlaid bladder
2. bulb and valve
3. stethoscope
4. soft bag
Technical alterations reserved.