20
dent investigators using the BHS protocol. This is the highest grading avail-
able for blood pressure monitors. Please see bhsoc.org for more information.
13. Technical specifications:
Weight: 148g (with batteries + cuff)
Size: 85 x 77 x 75mm (including cuff)
Storage temperature: –20 to +55°C (-4°F to +131°F)
Humidity: 15 to 90% relative humidity
maximum
Operation temperature: 10 to 40°C (50°F - 104°F)
Display: LCD-Display (Liquid Crystal
Display)
Measuring method: oscillometric
Pressure sensor: capacitive
Measuring range: SYS/DIA: 30 to 280 mmHg
Pulse: 40 to 200 per minute
Cuff pressure display range: 0-299 mmHg
Memory: Stores the last 99 measurements
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg or 2% of
reading >200mmHg
Pulse ± 5 % of the reading
Power source: 2 dry cells (Batteries)
size AAA 1.5V
Accessories: storage case
Technical alterations reserved!