3. Choosing between 3 different Nebulisation Rates
The nebuliser generates an aerosol with variable characteristics to
adapt to the patient’s respiratory pathology. It can produce an
aerosol with the granulometric characteristics that will deposit the
maximum amount of the drug in the respiratory tract being treated.
Choose one of the 3 following therapeutic treatment settings by
regulating the wheel on the nebuliser’s 5 upper body:
4. Cleaning and Disinfecting
Thoroughly clean all components to remove medication residuals
and possible impurities after each treatment.
The compressor 1 and the air hose 6 should be cleaned with a
clean, moist cloth.
Cleaning with water
Disinfecting
Sterilizing with steam
5. Maintenance, Care, and Service
Order all spare parts from your dealer or pharmacist, or contact
Microlife-Service (see foreword).
6. Malfunctions and Actions to take
The device cannot be switched on
The nebuliser functions poorly or not at all
7. Guarantee
This device is covered by a 2 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
8. Technical Specifications
7.
Empty the remaining medication from the nebuliser and clean the
device as described in the section «Cleaning and Disinfecting».
•
Upper air passages (nose, throat, etc.): Position the wheel
towards the bottom at number 3; the aerodynamic diameter of
the aerosol particles will be from 10 to 14 micron.
•
Tracheal / bronchial tract: Position the wheel in the center
at number 2; the aerodynamic diameter of the aerosol particles
will be from 6 to 9 micron.
•
Deep air passages: Position the wheel towards the top, at
number 1; the aerodynamic diameter of the aerosol particles
will be from 1 to 5 micron.
)
Always wash your hands well before cleaning and disin-
fecting the accessories.
)
Do not expose the compressor to water or heat.
)
Replace the air hose for each treatment with a new patient
or in case of impurities.
Unplug the device before cleaning.
X
Wash all nebuliser components (except air hose) under
warm tap water (max 60 °C) for about 5 minutes adding if
necessary a small quantity of detergent following dosage and
use limitations as provided by detergent manufacturer.
X
Rinse thoroughly making sure that all deposits are washed
away and leave to dry.
X
All nebuliser components (except air hose) can be disinfected
with chemical disinfectants following dosage and use limita-
tions as provided by disinfectant manufacturer.
X
Disinfectants are usually available at pharmacies.
X
All nebuliser components (except air hose and masks) can be
heat steam sterilized up to 121 °C (20 min.) or 134 °C (7 min.).
EN554/ISO11134.
X
The sterilisation packaging must conform to EN868/ISO11607
and be suitable for steam sterilisation.
X
After sterilization always let all components cool down to
ambient temperature before further use.
)
Do not repeat sterilization cycle when components are still
warm.
X
We recommend replacing the nebuliser, mouthpiece, and the
face masks after 3 months use.
X
Check the filter continually for cleanliness and replace it if dirty,
or after a maximum of 3 months use.
Spare filters are provided with the device.
X
To replace the filter, open the air filter compartment 4 on the
top of the device marked «Filter» and exchange the filter AT.
•
Ensure the power cord 2 is correctly plugged into the socket.
•
Ensure the ON/OFF switch 3 is in the position «I».
•
The device has automatically switched off due to overheating.
Wait until it has cooled down and try again.
•
Ensure the air hose 6 is correctly connected at both ends.
•
Ensure the air hose is not squashed, bent, dirty or blocked. If
necessary, replace with a new one.
•
Ensure the nebuliser 5 is correctly assembled AK.
•
Ensure the required medication has been added.
•
The guarantee covers only the compressor. The replaceable
components like nebuliser, masks, mouthpiece, air hose, and
filters are not included.
•
Opening or altering the device invalidates the guarantee.
•
The guarantee does not cover damage caused by improper hand-
ling, accidents or non-compliance with the operating instructions.
Nebulisation rate: Pos. 1: 0.23 ml/min
Pos. 2: 0.52 ml/min
Pos. 3: 0.68 ml/min
Average fill volume: 4 ml NaCl 0.9%
Particle size: Pos. 1: 1.82 μm* / 4.59 μm**
Pos. 2: 2.97 μm* / 7.47 μm**
Pos. 3: 4.03 μm* / 10.13 μm**
* API system / ** Malvern system
Compressor air flow: 15 l/min.
Acoustic noise level: 53 dBA
Power source: 230V~ / 50Hz / 150VA
Power cord length: 1.5 m
Nebuliser capacity: min. 2 ml; max. 12 ml
Residual volume: 0.7 ml
Operating temperature: 10 - 40 °C / 50 - 104 °F
10 - 95 % relative maximum humidity