Type: Fingertip Pulse Oximeter
OXY 300
Display: OLED display
SpO
2
:
Measurement range: 70 ~ 100 %
Accuracy: 70 ~ 100 %: ±2 %
Resolution: 1 %
Pulse rate:
Measurement range: 30 ~ 250 bpm
Accuracy: 30 ~ 99 bpm: ±2 bpm
100 ~ 250 bpm: ±2 bpm
Resolution: 1 bpm
Operating
conditions:
5 - 40 °C / 41 - 104 °F
≤ 80 % relative maximum
humidity
Storage conditions: -20 - +55°C / -4 - +131 °F
≤ 93 % relative maximum
humidity
Automatic switch-off: Automatically shut down
in 8 seconds, when no or
low signal is detected.
Battery: 2 x 1.5 V alkaline
batteries; size AAA
Battery lifetime: approx. 30 hours (using
new batteries)
Weight: 56 g (including batteries)
Dimensions: 58 x 32 x 34 mm
IP Class: IPX1
Reference to
standards:
EN ISO10993-1/-5/-10;
IEC 60601-1; EN 60601-
1-2; EN ISO9919; EN
62304; EN 60601-1-6;
CE0123
Expected service life: 5 years (when used 15
times/day; 20 minutes for
each measurement)