EasyManua.ls Logo

Mikro MK201A - Technical Data and Indicators; Current and Time Adjustments; Light Indicator Status

Mikro MK201A
4 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TECHNICAL DATA
1. Current and Time Adjustments
Earth-fault Low-set Current (l>) Adjustment
• This adjustment is for setting the minimum earth-fault current for tripping. If the
measured current exceeds this threshold value, tripping starts. After a prefixed
delay time determined by the DELAY adjustment, the trip contacts will be activated.
• The setting range is from 0.1A to 2A.
Earth-fault High-set Current (l>>) Adjustment
• This adjustment is for setting the instantaneous tripping current due to an earth-fault.
• The setting range is from 1x to 10x of the earth-fault low-set setting value.
l>> = a x l>, a = 1 to 10
• This high-set feature can be disabled by setting the tripping current to infinity ( ' )
Time Delay (DELAY) Adjustment
• This time delay setting is for setting the delay time from the start of the earth-fault
low-set tripping to actual tripping of the relay contact.
• The setting range is from 0.05 sec to 1.0 sec.
2. Light Indicators
The light indicators display the status of the system.
Indicator
AUX l> l>> Status
Off Off Off No auxiliary power supply.
On Off Off System normal mode. No tripping.
On On Off
Earth-fault low-set start.
On
Blink
Off
Earth-fault low-set tripped.
On Off On
Earth-fault high-set start.
On Off Blink Earth-fault high-set tripped.

Related product manuals