Mahr GmbH, Millimar C1208/C1216/1240
18 Technical data
General
Dimensions (Height x width x depth) 205 x 160 x 165 mm
Weight 2.1 kg
Power supply 9 V DC
through a wide-range power supply unit 100 V up to 240 V, 47 Hz up to 63 Hz
Power consumption 10 W
System of protection IP54 (IP43 for conductive dust), IEC60529
Carrier frequency
C 1208 M / C 1216M / C 1240M 20 kHz
C 1208 T / C 1216T 13 kHz
C 1208 F / C 1216F 5 kHz
Magnitude of excitation voltage
C 1208 M / C 1216M / C 1240M 5 V
C 1208 T / C 1216T 3 V
C 1208 F / C 1216F 2 V
Measuring range
C 1208 4000 µm (+/- 2000 µm)
C 1216 / C 1240 4000 µm (+/- 2000 µm)
400 µm (+/- 200 µm)
Sensitivity adjustment range
max. deviation for chan. 2 referred to chan. 1 +/- 1.2 %
Resolution (max.)
C 1208 0.1 µm
C 1216 / C 1240 0.01 µm
Cut-off frequency 60 Hz (-3dB); adjustable via „FILTER (4.2)
Response time
Depends on the value adjusted for „FILTER (4.2)“.
As a rule, the shortest possible response time is detailed:
Analog scale 100 ms
Numerical display 100 ms
Analog output 20 ms
Digital output 20 ms
Error limits
Scale division of ± 10 2.5 %
Scale division of ± 3 2 %
Numerical display 0.2 µm or 0.3 % of the indicated probe value*
Analog output 5 mV or 0.5 % of the output voltage*
Digital output 0.2 µm or 0.3 % of the indicated probe value*
* As a rule, the larger of the two values in question is valid.