EasyManua.ls Logo

Millipore WP6111560 - User Manual

Millipore WP6111560
7 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
SigmaAldrich.com
PR036484w REV 07/20 1 of 7
User Guide
Chemical Duty
Vacuum/Pressure Pump
WP6111560
WP6122050
WP6110060
Introduction
The vacuum/pressure pump is a continuously running
constant air flow type unit for use with laboratory
equipment. This vacuum/pressure pump combines
all the advantages of lower pressure pumps with the
perfect uniformity of diaphragm type operation. It
compresses air by means of a long-life diaphragm,
which flexes within the compressor. This pump is
intended for filtration of liquids or gases, or for
other continuous or intermittent use, with all types
of filter holders.
CAUTION: Always use the supplied hydrophobic vent
filter or a vacuum-flask water trap in conjunction with
the pump (see Figures 1 and 2). Never pump or draw
liquids through the pump, as this will damage it.
The pump has these features:
The motor and pump are permanently lubricated.
A thermal overload switch with automatic reset
protects the motor.
Low operating noise is further minimized
by rubber feet.
All internal surfaces (except the stainless steel
leaf valves) are coated with PTFE polymer, to
prevent corrosion.
The diaphragm is PTFE-lined to prevent attack by
chemicals or solvent vapors.
A Millex
®
-FA
50
hydrophobic vent filter and
27 in. (69 cm) of
1
4 in. (6 mm) I.D. silicone
vacuum tubing are supplied with the pump.
Rules for Safe Operation
Y
The motor is thermally protected and will
automatically restart unexpectedly when
the overload device resets.
Do NOT pump flammable or explosive gases or
vapors or operate this pump in an atmosphere
containing flammable or explosive gases
or vapors.
Use of this pump in a manner not specifically
stated in this user guide may result in severe
bodily injury.
Read and understand the information in this owner’s
manual before operating the vacuum/pressure pump.
The vacuum/pressure pump should be operated in a
dry, clean, and well ventilated area.
NOTE: To ensure proper cooling, the
vacuum/pressure pump cooling fan must not be
blocked and should be a minimum of 1 in. (2.5 cm)
away from any wall or obstruction.
When the unit is not in use, wrap the power cord
around the vacuum/pressure pump and store in a
dry place. Do not abuse the cord.
Replace the Millex
®
-FA
50
vent filter (SLFA05010)
if necessary.
Inspect hose, plug, and cord for signs of damage
before use. Do not use if a deficiency is found.
Never operate a damaged unit. For replacement
parts, see Technical Assistance section for details.
This vacuum/pressure pump needs no lubrication.
Applying oil to any part may result in polluted air
delivery to the air-handling equipment and will
damage the pump.
To operate at maximum efficiency, the pump system
must be thoroughly clean. Refer to the Maintenance
section for details on properly cleaning the pump.
Compressed air blast must never be aimed
at anyone because it can cause serious injury.
Keep children away.
Question and Answer IconNeed help?

Do you have a question about the Millipore WP6111560 and is the answer not in the manual?

Summary

Product Introduction

Safety Guidelines and Precautions

Pump Usage Guidelines

Labeling Symbols Overview

Operating the Vacuum;Pressure Pump

General Maintenance Information

Pump Component Replacement

Diaphragm Replacement Procedure

Step-by-step guide for replacing the pump's diaphragm and liner.

O-ring Replacement Procedure

Instructions for removing and replacing the O-ring gasket in the pump head.

Troubleshooting Common Issues

Troubleshooting Vacuum Issues

Identifies causes and solutions for poor vacuum issues in the pump system.

Troubleshooting Pressure Issues

Details common causes and fixes for problems related to pump pressure.

Pump Repair and Parts

Replacement Part Kits Overview

Information on available part kits and recommended repair procedures.

Specifications, Ordering, and Support

Technical Product Specifications

Technical details including performance, dimensions, materials, and environmental conditions.

Ordering and Catalogue Numbers

Catalogue numbers and ordering details for pumps and replacement parts.

Product Notices and Support

Includes notice, contact info, technical assistance, and warranty details.

Summary

Product Introduction

Safety Guidelines and Precautions

Pump Usage Guidelines

Labeling Symbols Overview

Operating the Vacuum;Pressure Pump

General Maintenance Information

Pump Component Replacement

Diaphragm Replacement Procedure

Step-by-step guide for replacing the pump's diaphragm and liner.

O-ring Replacement Procedure

Instructions for removing and replacing the O-ring gasket in the pump head.

Troubleshooting Common Issues

Troubleshooting Vacuum Issues

Identifies causes and solutions for poor vacuum issues in the pump system.

Troubleshooting Pressure Issues

Details common causes and fixes for problems related to pump pressure.

Pump Repair and Parts

Replacement Part Kits Overview

Information on available part kits and recommended repair procedures.

Specifications, Ordering, and Support

Technical Product Specifications

Technical details including performance, dimensions, materials, and environmental conditions.

Ordering and Catalogue Numbers

Catalogue numbers and ordering details for pumps and replacement parts.

Product Notices and Support

Includes notice, contact info, technical assistance, and warranty details.

Overview

The Chemical Duty Vacuum/Pressure Pump is a continuously running, constant air flow unit designed for use with laboratory equipment. It operates by compressing air using a long-life diaphragm, which flexes within the compressor, providing the uniform operation characteristic of diaphragm-type pumps. This pump is suitable for various applications, including the filtration of liquids or gases, and other continuous or intermittent laboratory uses with all types of filter holders.

Function Description

The pump is engineered to provide both vacuum and pressure capabilities. It features a pump head, a vacuum gauge, and a pressure gauge to monitor and control the desired vacuum or pressure levels. The unit is equipped with a vacuum valve and a pressure valve, allowing users to adjust the output precisely. An on/off switch located on the front of the pump controls its operation.

For vacuum applications, the vacuum is adjusted by closing the pressure regulator and fully opening the vacuum regulator (counterclockwise). The vacuum regulator is then slowly tightened clockwise until the desired vacuum reading is achieved on the vacuum gauge. Conversely, for pressure applications, the pressure is adjusted by turning the vacuum regulator fully down and opening the pressure fully (counterclockwise). The pressure regulator is then slowly tightened clockwise until the desired pressure reading is obtained on the pressure gauge.

Usage Features

Safety is a paramount consideration in the design and recommended usage of this pump. It is crucial to always use the supplied hydrophobic vent filter or a vacuum-flask water trap in conjunction with the pump to prevent liquids or vapors from entering and damaging it. The pump should never be used to pump or draw liquids directly. For maximum protection against vapors and gases, using both the Millex®-FA50 vent filter and a vacuum trap is recommended. The filter flask and vacuum trap should be emptied after each use.

The pump is designed for indoor use in a dry, clean, and well-ventilated area. To ensure proper cooling and prevent overheating, the pump's cooling fan must not be blocked and should be positioned at least 1 inch (2.5 cm) away from any wall or obstruction. The unit rests on four vibration isolator pads, which help minimize operating noise.

Before operation, users should unpack the unit and retain all packing material until proper product operation is verified. The pump should be placed on a suitable surface, such as a bench, desk, or table, ensuring that ventilation holes on the motor housing are not blocked. Appropriately sized tubing, typically 1/4 inch I.D., should be selected to withstand the anticipated pressure or vacuum.

When setting up, prepare the necessary equipment or filter holder and connect the tubing to the equipment first, but do not connect the tubing to the pump yet. After plugging the power cord into an appropriate electrical source, turn on the pump using the toggle switch. Only after the pump is operating should the tubing be connected from the filter holder or other equipment to the pump. If the pump is inadvertently turned on with the tubing fully connected and does not run, it should be turned off, the tubing disconnected, and then restarted. If the thermal overload switch activates, disconnect the tubing and allow the pump to cool for at least 10 minutes before restarting.

The pump is supplied with a Millex®-FA50 hydrophobic vent filter and 27 inches (69 cm) of 1/4 inch (6 mm) I.D. silicone vacuum tubing. The motor and pump are permanently lubricated, and a thermal overload switch with automatic reset protects the motor. All internal surfaces, except the stainless steel leaf valves, are coated with PTFE polymer to prevent corrosion, and the diaphragm is PTFE-lined to resist attack by chemicals or solvent vapors.

Maintenance Features

The chemical duty vacuum/pressure pump is designed for many hours of trouble-free operation under normal conditions and with proper handling. It is a 100% oil-free dry vacuum/pressure pump, employing a non-lube piston and cylinder. The bearings are sealed and permanently lubricated, requiring no maintenance or lubrication attempts. The units are built for continuous duty operation, emphasizing quietness and durability.

It is critical not to lubricate any parts with oil, grease, or petroleum products, as this can lead to polluted air delivery and damage the pump. Cleaning should not be done with acids, caustics, or chlorinated solvents. The connecting rod or motor bearings should not be replaced.

Before performing any maintenance, the pump must be disconnected from the power source.

Regular maintenance includes replacing the Millex®-FA50 vent filter (SLFA05010) when necessary. The pump system must be thoroughly clean to operate at maximum efficiency.

For diaphragm replacement:

  1. Unscrew the four socket-cap screws from the pump head and the two screws from the pump handle. Remove the pump head assembly.
  2. Unscrew the four Phillips-head screws, then remove the retainer plate, diaphragm liner, and diaphragm.
  3. Replace the diaphragm, ensuring the printed number faces up and aligning holes with screw holes in the pump top.
  4. Replace the diaphragm liner, aligning its holes with the screw holes.
  5. Replace the retainer plate, aligning the four small holes with the screw holes and orienting the large hole to the left as shown. Tighten the Phillips-head screws, taking care not to nick the screw heads, as burrs can damage the inlet valve.
  6. Replace the pump head assembly and tighten the six screws.

For O-ring replacement:

  1. Unscrew the four socket-cap screws from the pump head and the two screws from the pump handle. Remove the pump head assembly.
  2. Remove the five Phillips-head screws holding the valve plate assembly to the underside of the pump head assembly.
  3. Remove the O-ring and replace it by pressing the new O-ring into the groove.
  4. Replace the valve plate assembly with the recessed screw holes facing out and tighten the five Phillips-head screws.
  5. Replace the pump head assembly and tighten the socket head screws.

In case of troubleshooting, issues like low flow, low pressure, the unit not starting, motor overheating, or loud operation can be addressed. Common causes include damaged valves, debris in valves, worn diaphragm, loose head screws, bent motor shaft, loose fittings, insufficient ventilation, loose rod clamping screw, and damaged O-rings. Corrective actions range from replacing flapper valves, removing debris, replacing the diaphragm and/or liner, tightening screws, replacing the entire unit, increasing air circulation, applying Loctite® 242, and replacing O-rings.

Two replacement part kits are available: a Pump Maintenance Kit (WP61MNT00) and a Pump Rebuild Kit (WP61RBD00). The maintenance kit includes a diaphragm, diaphragm liner, O-ring gasket, head screws, and handle screws. The rebuild kit includes a hold-down plate assembly, valve plate assembly, pump head, vacuum gauge, pressure gauge, head screws, and handle screws. It is recommended that trained service technicians perform repairs.

Millipore WP6111560 Specifications

WP6111560

Electrical Specifications IconElectrical Specifications
voltage115 V~, 60 Hz
current3.5 A
motor power1/15 HP
Environmental Conditions IconEnvironmental Conditions
temperature range5 °C (41 °F) to 40 °C (104 °F)
relative humidity80% for temperatures up to 31 °C, Max. decreasing to 50% at 40 °C
altitudeMax. 2, 000 meters
Weight and Dimensions IconWeight and Dimensions
length7.5 in. (19.1 cm)
width7 in. (17.8 cm)
height8 in. (20.3 cm)
weight9 lbs (4.1 kg)

WP6110060

Electrical Specifications IconElectrical Specifications
voltage100 V~, 50/60 Hz
current4.0 A
motor power1/15 HP
Environmental Conditions IconEnvironmental Conditions
temperature range5 °C (41 °F) to 40 °C (104 °F)
relative humidity80% for temperatures up to 31 °C, Max. decreasing to 50% at 40 °C
altitudeMax. 2, 000 meters
Weight and Dimensions IconWeight and Dimensions
length7.5 in. (19.1 cm)
width7 in. (17.8 cm)
height8 in. (20.3 cm)
weight9 lbs (4.1 kg)

WP6122050

Electrical Specifications IconElectrical Specifications
voltage230 V~, 50 Hz
current1.7/1.1 A (Schuko) plug
motor power1/15 HP
Environmental Conditions IconEnvironmental Conditions
temperature range5 °C (41 °F) to 40 °C (104 °F)
relative humidity80% for temperatures up to 31 °C, Max. decreasing to 50% at 40 °C
altitudeMax. 2, 000 meters
Weight and Dimensions IconWeight and Dimensions
length10.5 in. (26.7 cm)
width9.4 in. (23.9 cm)
height9.7 in. (24.6 cm)
weight12 lbs (5.4 kg)

Related product manuals