E - 2     
The equipment 
delivers a shock.
Shock advised. Everyone clear. Notifies a shockable rhythm has 
been detected.
Shock will be delivered in: 3, 2, 1 Prompts the equipment is fully 
charged and is preparing to deliver 
a defibrillation shock.
Shock delivered. Prompts the shock is delivered.
Press flashing shock button Prompts the equipment is fully 
charged and ready to deliver the 
defibrillation shock. 
Shock canceled. Shock button was not 
pressed. 
The Shock button is not pressed 
within 30s and the equipment 
cancels the shock.
Device error, charge failed. The equipment is unable to start 
charging because of a fault 
condition. The equipment resumes 
the rhythm analysis after a charging 
failure. After three consecutive 
charging failures, the equipment 
automatically enters the CPR status.
Device error, shock failed. The equipment is unable to deliver a 
shock because of a fault condition. 
Or, it is not suitable to deliver a 
shock to the patient. The equipment 
disarms itself and resumes the 
rhythm analysis after a discharging 
failure. After three consecutive 
discharging failures, the equipment 
automatically enters the CPR status.
Shock canceled. Press pads firmly to 
patient's bare skin.
Shock canceled. Pads must not be 
touching each other.
Rhythm change, shock canceled The equipment detects a rhythm 
change and cancels the shock
Perform CPR Start CPR immediately. Prompts to prepare to provide 
compressions and breaths CPR.
Give chest compressions immediately. Prompts to prepare to provide 
compressions-only CPR.
Continue to compress without rescue 
breaths. 
Place one hand on center of chest, the 
other hand should be on top of first hand. 
Interlock the fingers. Continue to push 
down hard.
Place one hand on center of chest. Keep 
arms straight. Continue to push down 
hard.
Keep arms straight.Continue to push 
down hard.
Interlock the fingers. Continue to push 
down hard.
100 compressions remaining.
50 compressions remaining.
20 compressions remaining.
Condition Voice Prompt Description