A - 9
360 J defibrillation waveform into impedance of 25Ω, 50Ω, 75Ω, 100Ω, 125Ω, 150Ω, 175Ω
Internal defibrillation 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, 50 J
Patient impedance range
External defibrillation 25 to 300 
Internal defibrillation 15 to 300 
Synchronized discharge delay
Local synchronized discharge delay < 60ms (from the peak of R-wave)
Remote synchronized discharge delay < 25ms (from the rise edge of synchronous signal)
AED
Shock series Energy level: 100 to 360J, configurable for adult;
10 to 100J, configurable for pediatric
Shocks: 1, 2, 3, configurable;
Meeting AHA/ERC guidelines 2015 by default
AED ECG analysis performance Refer to CMindray Shockable Rhythm Analysis Algorithm.
Time (ms)
Voltage (V)