EasyManua.ls Logo

MiniMed 670G - Connect Reservoir to Tubing; Connect Tubing to Reservoir; Address Air Bubbles in Tubing; Remove Plunger from Reservoir

Default Icon
4 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
MINIMED™ 670G SYSTEM | MEDTRONIC EXTENDED INFUSION SET
CHANGING THE RESERVOIR ONLY
Now you will place the tubing connector onto the lled reservoir.
CONNECT THE RESERVOIR TO THE INFUSION SET TUBING
3
Select Next.
2
Turn the reservoir clockwise until you feel it
lock into place.
1
Place the reservoir into the pump.
Next, you will place the lled reservoir into the reservoir compartment on the pump.
PLACE THE RESERVOIR INTO THE PUMP
2
If you see air bubbles, tap reservoir to move them to
top. Push plunger just a bit to move them into tubing.
1
Gently push the tubing connector onto the reservoir.
Turn the connector clockwise until locked. You will
hear a click.
Connector
Plunger
3
Twist plunger counter-clockwise to
loosen and remove.
The backlight may have
turned o.
Press any button to turn
the screen back on.
Select Next.
Press to open the menu. If the pump is
locked, you will need to unlock the pump
after pressing
.
Select Load Reservoir from the
menu.

Other manuals for MiniMed 670G

Related product manuals