CHANGING THE RESERVOIR ONLY
MINIMED™ 780G SYSTEM | MEDTRONIC EXTENDED INFUSION SET
- 1 -
START HERE
1
Wash your hands.
2 3
Select New Reservoir.
4
Disconnect the tubing from the cannula housing. Do not
remove the cannula housing from your body.
6
With the infusion set tubing removed from your body and the reservoir
removed from the pump, select Rewind.
5
Remove the used reservoir from the pump.
Continue with “Fill the
Reservoir” on the next page.
NOTE: The Extended Wear infusion set can be used
for up to seven days. The reservoir can be used for
up to three days or seven days; refer to your reservoir
instruction for use. To avoid wasting insulin, ll the
reservoir with enough insulin to last until the next
infusion set change.
When you change the infusion set, change all of its
pieces (the cannula housing, tubing, and reservoir).
Disconnect the infusion set tubing connector from the used reservoir. Place the tubing
on a clean surface. Do not throw it away.
Dispose of the used reservoir in an appropriate container
according to local laws.
Select .

Other manuals for MiniMed 780G

Question and Answer IconNeed help?

Do you have a question about the MiniMed 780G and is the answer not in the manual?

MiniMed 780G Specifications

General IconGeneral
Device TypeInsulin Pump
ManufacturerMedtronic
ModelMiniMed 780G
Automated Insulin DeliveryYes
Continuous Glucose MonitoringYes
Bluetooth ConnectivityYes
SmartGuard TechnologyYes
CGM CompatibilityGuardian Sensor 3
Compatible Glucose MonitorGuardian Connect
Infusion SetCompatible with Medtronic infusion sets
Insulin DeliveryBasal and bolus
Battery LifeUp to 7 days
Reservoir Capacity3.0 mL (300 units)
Water ResistanceIPX8 rated (waterproof up to 3.6 meters for 24 hours)

Related product manuals