13
Command UCW
Function @Receives the standard value for measuring user calibration
standard and memorizes in the main body.
@
Calculate the coefficient for user calibration standard and memorizes.
(On this product side, the fourth digit is rounded off
because the display valid number of digit is 3-digit.)
@The unit is recognized with the unit set on this product side.
@
Resulting in error when the command is sent in the following condition.
Ë In case not to be calculated and memorizes normally
Condition that the measured value is not held in this product.
Condition that the measured value is the measuring range over.
Condition that the measured value is the display range over.
Condition that the measured value Y is "0".
Condition that the luminance standard value Y is "0".
Condition that the chromaticity standard value y is "0".
Control side Ë this product format
"UCW, ,D ,D "
This product Ë control side format
@In case to be calculated and memorized normally
"OK00"
@In case not to be calculated and memorizes normally
"ER01"
Control sideËthis product format details
@Following the command, outputs luminance standard value Y,
chromaticity standard value x, chromaticity standard value y.
This productËcontrol side format details
@In case to be calculated and memorized normally
Outputs the error check command only.
@In case not to be calculated and memorizes normally
Outputs the error check command "ER01".
Command CLE
Function @Carries out memory clear.
Clear of the color difference standard color data, the
arbitrary calibration standard data.
Control side Ë this product format
"CLE"
This product Ë control side format
"OK00"
Control sideËthis product format details
Outputs command only
This productËcontrol side format details
Outputs error check command only.