M9111
PC
BOARD
adjustment
a
ㄴㄴ
n-
cs
u
?
RE
2
ps
Le
mel
aa
|
m,
|
|
9
Lo
c
Q
gayi
e
L
el
TF
=
This
adjustment
has
to
be
done
after
connect
all
of
connectors
completely
But
actual
exposure
is
not
necessary.
*
Each
PC
boards
delivered
from
Mikasa
Factory
have
already
been
adjusted.
1.
Adjustment
of
VR3
(
DC370Y
)
Adjust
voltage
between
370V
(+)
and
GND
(-)
to
be
368.5
-
372V
by
turning
VR3.
Voltage
is
increased
by
turning
clockwise.
|
2.
Adjustment
of
VR2
(
Over
Current
Protection
circuit)
Adjust
voltage
between
OCP
(+)
and
GND
(-)
to be
1.09
-
1.11Y
by
turning
VR2.
[Voltage
is
increased
by
turning
clockwise.
|
3.
Adjustment
of
VRI
(
Over
Voltage
Protection
circuit
)
Adjust
voltage
between
OVP
(+)
and
GND
(=)
to
be
@V
by
turning
VRI.
Woltage
is
decreased
by
turning
clockwise.
|
If
actual
measured
voltage
would
be
B,
a
can
be
calculated
as
following.
a=B
x
0.
0061