7.
Readjustment
of
exposure
time
This
adjustment
has
to
be
done
after
connect
all
of
connectors
completely.
Actual
exposure
is
necessary.
Note
1:
This
adjustment
requires
that
an
exposure
be
made.
Please
observe
all
radiation
related
safety
precautions.
Check
the
exposure
time
by
using
an
external
exposure
time
meter
such
as
the
VICTOREEN
NERO.
X-ray
unit
setting:
82kV
-
0.5
sec.
Adiustment:
Exposure
time
is
decreased
by
turning
clockwise.
(VR6
on
the
M3108A
PC
board)
-
16
-