Details about power requirements, voltage range, frequency, and consumption.
Information on the inverter and filament circuit specifications.
Specifies the warm-up duration after power on and exposure.
Technical details on tube voltage, kV steps, and tube current.
Specifications for the TOSHIBA D-124S X-ray tube.
Details on collimator features and filtration materials.
Includes weight, size, power cord, and exposure cord lengths.
Steps for connecting power, turning on, and initial warm-up.
Procedures for adjusting kilovoltage and exposure time.
How to use the collimator to set the radiation field.
Step-by-step guide for radiography, including safety checks.
Explains operational indicators and warning lamps like WARM UP and ERROR.
Notes that PC boards are factory adjusted; details on 24VDC supply.
Procedures for adjusting kV and mA using the M3108A board.
Method for adjusting exposure time with an oscilloscope.
Critical safety warnings for board adjustments and connector handling.
Explanation of short pins (LBD, 10SEC, TEST) on the M3108A board.
Details adjustments for over voltage, over current, and DC voltage circuits on M9111.
Lists adjustments for frequency, pre-heat time, kV, and mA on M9112.
Notes that exposure is required and safety precautions must be observed.
Lists necessary equipment and oscilloscope settings for kV adjustment.
Detailed steps for adjusting kV based on EP waveform at 70kV and 100kV.
Warning about external kV tester deviation and recommended adjustment range.
Notes that exposure is required and safety precautions must be observed.
Lists necessary equipment and oscilloscope settings for adjustment.
Detailed steps for adjusting pre-heat time using VR2-VR3 with acceptable limits.
Procedure to adjust mA by turning VR5 to achieve a specific IP waveform.
States that this adjustment requires an exposure and safety precautions.
Recommends using an external exposure time meter for checking.
How to adjust exposure time using VR6 on the M3108A PC board.
Displays sample output waveforms for 100kV and 20mA.
Schematic diagram for the M9111-1 board.
Schematic diagram for the M9112 board.
Part of the schematic diagram for the M3107A board.
Another part of the schematic diagram for the M3107A board.
Shows an exploded view with numbered parts for assembly.
Lists parts for HF100H with index, part number, and designation.
Lists parts for HF100 with index, part number, and designation.
| Type | Portable X-ray generator |
|---|---|
| Voltage Range | 40 - 100 kV |
| Battery Life | Up to 100 exposures |
| Control | Digital |
| Output | X-ray |
| kV Range | 40 - 100 kV |
| Portability | Portable |
| kVp Range | 40 - 100 kVp |
| mAs Range | 0.1 - 100 mAs |
| Generator Type | High frequency |
| Cooling | Air-cooled |
| Display | Digital |