4. Specications
Test Voltage 500V, 1000V, 2500V, 5000V
Insulation resistance
1TΩ / 500V
2TΩ / 1000V
5TΩ / 2500V
10TΩ / 5000V
Accuracy
0~100GΩ / 500V
0~200GΩ / 1000V
0~500GΩ / 2500V
0~1000GΩ / 5000V
±(5.0%rdg + 5dgt)
100G~1TΩ / 500V
200G~2TΩ / 1000V
500G~5TΩ / 2500V
1000G~10TΩ / 5000V
±12%rdg
Resolution
1000MΩ: 1MΩ
10GΩ: 0.01GΩ
100GΩ: 0.1GΩ
1TΩ: 1GΩ
10TΩ: 10GΩ
Short circuit current up to 5mA
PI (Polarization Index)
DAR (Dielectric Absorption Ratio)
Voltmeter
ACV: 30~600V
DCV: 30~600V
Accuracy: ±(2.0%rdg + 3dgt)
Resolution: 1V
Current measurement
0.5nA ~ 0.55mA
(Depending on the insulation resistance)
Power source Rechargeable battery
Adapter
Input: 100-240Vac, 0.4A, 50-60Hz
Output: 24Vdc, 0.62A
Dimensions 330(L) × 260(W) × 160(D)mm
Weight Approx. 3760g
Safety standard
EN 61010-1 CAT IV 600V
EN 61010-2-030 EN 61326-1
Accessories
Instruction manual Test leads
Data transmission cable CA-232
Compact disk (CD) for PC interface
Alligator clip Charger Test report