Blood pressure classication in adults
6.10. Description of technical alarms
The monitor will display “Hi” or “Lo” as a technical alarm on the screen without delay if the
determined blood pressure (systolic or diastolic) is outside the nominal range specied in the
SPECIFICATIONS section. In this case, you should consult a physician or check if the operation
complies with the instructions.
The technical alarm condition (out of nominal range) is set at the factory and cannot be adjusted
or disabled. This alarm condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm does not latch and no reset is required. The signal displayed on the LCD
will disappear automatically after approximately 8 seconds.
Normal
Normal-high
Mild hypertension
Moderate hypertension
Severe hypertension
Systolic
(mmHg)
Diastolic
(mmHg)
Classication
Pressure
PSS (mmHg) PSD
(mmHg)
Colour
indicator
Optimal <120 <80 green
Normal 120-129 80-84 green
Normal-high 130-139 85-89 green
Hypertension
grade 1
140-159 90-99 yellow
Hypertension
grade 2
160-179 100-109 orange
Hypertension
grade 3
≥180 ≥110 red
WHO denitions and classication of blood
pressure levels