EasyManua.ls Logo

Monaghan Aerobika - Intended Use and Safety; Intended Use; Cautions; Contraindications

Default Icon
9 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN EN
2 3
INTENDED USE
The Aerobika
®
Oscillating Positive Expiratory Pressure Therapy System is intended for use as a Positive Expiratory
Pressure (PEP) device. The Aerobika
®
Oscillating PEP device may also be used simultaneously with nebulized aerosol
drug delivery. The device is intended to be used by patients capable of generating exhalation flow of 10 lpm for 34
seconds.
CAUTIONS
• Donotleavethedeviceunattendedwithchildren.Itisnotatoy.
• Donotpressattachmenttabswhenthecaseisdisassembled.Doingsocouldbreakthetabsandpreventproper
assembly of the device.
• Ensurecontraindicationsareobserved.
• Federal(USA)lawrestrictsthesaleofthisdeviceonorbytheorderofaphysician.
CONTRAINDICATIONS
UseoftheAerobika
®
Oscillating PEP device is not recommended for the following conditions
• Patientsunabletotoleratetheincreasedworkofbreathing
• Intracranialpressure(ICP)>20mmHg
• Hemodynamicinstability(bloodpressureinstability)
• Recentfacial,oral,orskullsurgeryortrauma
• Acutesinusitis
• Epistaxis(bleedingnose)
• Esophagealsurgery
• Nausea
• Activehemoptysis(bleedingfromlungs)
• Untreatedpneumothorax(untreatedcollapsedlung)
• Knownorsuspectedtympanicmembraneruptureorothermiddleearpathology
WARNING
Useofthisdeviceatexcessivepressuresmayhaveadverseeffects.Expiratorypressuresabove20cmH
²
O in patients
sensitive to increased transpulmonary pressure may develop one or more of the adverse side effects listed below. Expert
clinicaljudgmentshouldbeexercisedintheselectionoftheappropriatesettingforeachindividualpatient.Failureto
match the appropriate resistance setting on the Adjustment Indicator with the patient’s expiratory flow may result in
failure to achieve therapeutic objectives of oscillatory PEP therapy or one or more adverse side effects listed below.
Adverse reactions may include
• Increasedworkofbreathingthatmayleadtohypoventilationandhypercarbia
• Increasedcranialpressure
• Cardiovascularcompromise
• Myocardialischemia
• Decreasedvenousreturn
• Airswallowingwithincreasedlikelihoodofvomitingandaspiration
• Claustrophobia
• Pulmonarybarotraumas
SETTING UP THE AEROBIKA
®
OSCILLATING PEP
FordevicedescriptionsseeFigure1.
Beforeuse,ensuretheseinstructionshavebeenreadandarekeptavailableatalltimes.Ifyouhavequestionsaboutthe
performanceorusabilityofthisproduct,pleasecontactyourhealthcareprofessionalpriortouse.
FIGURE 2
DETERMINING PROPER RESISTANCE SETTING
The Aerobika
®
Oscillating PEP device should be set up by a healthcare professional using the following steps:
1. StartwiththeResistanceIndicatoratthelowestsetting.(Fig.2)
2. With the patient in a comfortable seated position, have them
inhalethroughthedevicetakingadeeperbreaththannormal,
but not to total lung capacity, then exhale briskly but not
forcefully through the device.
3. Resistance should be set so exhalation lasts for 34 times
longer than inhalation. Resistance can be increased by moving
theindicatortotheleft.Toreduceresistance,movetheindicator
backtotheright.
NOTE: Selection of the proper resistance setting will produce
thedesiredinspiratory:expiratoryflowratioof1:3or1:4for10–20minuteswithoutexcessfatigue.
4. Patientsshouldbereevaluatedperiodicallyorwithanychangeinstatus.Asairwaysbecomelesscongested,more
resistance will be needed to maintain inspiratory:expiratory ratios.
CAUTION: The resistance setting should be adjusted as recommended by your healthcare professional.
PROPER TECHNIQUE FOR USING YOUR AEROBIKA
®
OSCILLATING PEP
Treatments should be conducted twice daily or as directed by your healthcare professional.
1 23
4
56
7
1. Before use carefully examine the device. Replace immediately if any defect is noticed. Always ensure the Resistance
Indicatorispositionedasprescribedbyyourhealthcareprofessional.Gentlypressthetwohalvestogethertomake
sure the Attachment Tabs are engaged.
2. TheAerobika
®
Oscillating PEP can be used in any orientation. It is recommended that treatments be done in a
comfortably seated position.
3. Putthemouthpieceinmouthandcloselipsaroundittoensureaneffectiveseal.Makesureyourhanddoesnot
blocktheexhalationpathonthebackofthedevice.
4. Inhalethroughthedevicetakingadeeperbreaththannormal,butdonottotallyfillyourlungs.Holdyourbreathfor
2–3secondsbeforeexhaling.
5. Exhale actively, but not forcefully, through the device. Ideally, exhalation should last 34 times longer than
inhalation.Keepyourcheeksflatandfirmtomaximizetreatmenteffectiveness.Ensureyoumaintainagoodsealon
the mouthpiece for the duration of your treatment.
Repeat
Steps
3–6
UNCONTROLLED WHEN PRINTED

Other manuals for Monaghan Aerobika

Related product manuals