4. TROUBLESHOOTING GUIDE
4. TROUBLESHOOTING GUIDE
NO FOOD Alarm
1. Is bag of pump de-
livery set empty?
YES
NO
2. Is air present in
the tubing?
NO
3. Is formula
blenderized or ag-
gressively mixed?
NO
4. Are air bubbles
trapped inside the
cassette?
YES
bag
IF NO:
IF YES:
YES
YES
NO
If alarm continues, contact your
healthcare provider or Moog Customer
Advocacy at (800) 970-2337 for assistance.
ERRA, ERRD, ERRF, ER01, ER02, or ER03 Alarms
1. Was pump door
open while pump
was turned on or
while running?
YES /
UNSURE
NO
If alarm continues, contact your
healthcare provider or Moog Customer
Advocacy at (800) 970-2337 for assistance.