4. TROUBLESHOOTING GUIDE
LOAD SET Alarm
1. Is door closed se-
curely?



2. Is door cracked
or tab on inside of
door broken?
YES
Contact your healthcare provider or Moog
Customer Advocacy at (800) 970-2337
to receive a new door.
NO /
UNSURE
NO
NO
Contact your healthcare provider or Moog
Customer Advocacy at (800) 970-2337
for assistance.
NO FLOW IN or NO FLOW OUT Alarms
4NO FLOW IN
4NO FLOW OUT
1. Is delivery set
tubing pinched,
kinked, or clogged?
YES
NO


2. Is tab on inside of
door broken?
YES
NO
Contact your healthcare provider or
Moog Customer Advocacy at (800) 970-2337
to receive a new door.
3. Is pressure sensor
region of cassette
receptacle clean?
NO
YES




4. Are there visible
signs of damage
to pressure sen-
sors area?
YES
Contact your healthcare provider or
Moog Customer Advocacy at (800) 970-2337
to return pump for service.
NO

If alarm continues, contact your
healthcare provider or Moog Customer Advocacy at
(800) 970-2337 for assistance.