24
ENGLISH
ELECTROMAGNETIC COMPATIBILITY DECLARATION
Guidance and manufacturer’s declaration- Electromagnetic emissions:
This device is intended for use in the electromagnetic environment specified below. The user of
this device should make sure it is used in such an environment.
EMISSION TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT- GUIDANCE
RF emissions CISPR 11 Group 1 The device uses RF Energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment
RF emissions CISPR11 Class B The device is suitable for use in all establishment.
Including domestic establishments and those di-
rectly connected to the public low-voltage power
supply network.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies
Guidance and manufacturer’s declaration - Electromagnetic immunity -
The Advance 1000/2000 devices are intended for use in the electromagnetic environment
specified below. The user of this device should make sure it is used in such an environment.
Immunity Test IEC60601
test level
Compliance Electromagnetic
Environment-Guidance
Electrostatic
Discharge (ESD)
IEC61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidi-
ty should be at least 30 %.
Electrical fast
transient/ burst
IEC61000-4-4
± 2kV for power
supply line
± 1kV for
input/out line
± 2kV for
power
supply line
± 1kV for
input/out line
Mains power quality should
be that of atypical
commercial or
hospital environment
Surge
IEC61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s)
to earth
± 1 kV line(s) to
line(s)
Mains power quality should
be that of atypical commer-
cial or hospital environment.
MA LTM500_3 02 A.indd 24 05/09/12 15:13