Firmly push down on the Glass Rim of the 1.0 mL BD
®
, pre- lled, Glass
Syringe so it is rmly set into the Syringe Housing. You may feel the syringe
“CLICK” into place. The Glass Rim should be touching the white collar.
DO NOT press on the Plunger Rod when inserting
syringe into Syringe Housing. This may result in
sterile seal being broken and loss of drug.
DO NOT remove the Orange Needle Cap Remover
until syringe is inserted and device reassembled.
4C
WARNING:
4A
4B
Step 4 – Insert 1.0 mL BD
®
, pre-fi lled, Glass Syringe into Syringe Housing
Remove the 1.0 mL BD
®
, pre- lled,
Glass Syringe from the protective
blister packaging. (1.0 mL BD
®
, pre-
lled, Glass Syringe is not included
with the WhisperJECT
™
Kit.)
DO NOT remove the 1.0 mL BD
®
,
pre- lled, Glass Syringe Needle Cap.
Place the Syringe Housing with
Orange Needle Cap Remover
onto a at surface, as shown.
Insert the syringe with the
Needle Cap end down into
the Syringe Housing.
16 17