Do you have a question about the MYLAN WhisperJECT and is the answer not in the manual?
Brand | MYLAN |
---|---|
Model | WhisperJECT |
Category | Medical Equipment |
Language | English |
Autoinjector lifespan is two years; store in protective case when not in use.
Once set up, the autoinjector enables subcutaneous injection.
Keep out of reach of children; do not insert other objects or twist depth past '10'.
Use only specified syringes, do not share, and do not replace cap remover after injection.
If the autoinjector is dropped, it must be replaced.
Refer to Patient Information Leaflet for support and safe syringe use.
Identifies components like Injection Button, Handle, Syringe Housing, and Viewing Window.
Illustrates syringe parts: Plunger Rod, Glass Rim, Needle Cap, and packaging.
Wash hands and gather autoinjector, syringe, sharps container, damp cloth, and alcohol swab.
Remove Handle, Orange Needle Cap Remover, and Syringe Housing from the protective case.
Insert the Orange Needle Cap Remover into the Syringe Housing; it does not click.
Place the Syringe Housing on a flat surface with the Orange Needle Cap Remover end up.
Position Handle over Syringe Housing and push down until a 'CLICK' is heard.
Verify reset by checking for no white check mark and a popped-up Orange Injection Button.
Rotate Orange Needle Cap Remover to align desired depth value with Needle Depth Indicator.
Smaller values (4, 6) give shallow injection; higher values (8, 10) give deeper injection.
Remove the 1.0 mL BD pre-filled Glass Syringe from its blister packaging. Do not remove needle cap.
Place Syringe Housing on surface and insert syringe with needle cap end down.
Firmly push Glass Rim into Syringe Housing until it 'CLICKS' into place.
Align white arrows and turn Syringe Housing until it locks into the Handle with a 'CLICK'.
Place autoinjector on clean surface, select site, and clean with alcohol swab.
Hold Handle and pull Orange Needle Cap Remover out of Syringe Housing to remove grey needle cap.
Turn Orange Needle Cap Remover upside down to allow both parts of the needle cap to fall out.
Save the Orange Needle Cap Remover for future use; do not dispose of it in the trash.
If needle cap is stuck, press on it until it dislodges.
Place autoinjector straight on site, press down until depth setting number is hidden.
Press Injection Button to start ('CLICK'), hold down, watch for green marks, then check mark.
Hold until check mark is visible or count 10 seconds after first 'CLICK' if window is not visible.
Confirm full dose delivery by checking the Viewing Window for a white check mark in a green box.
Contact your healthcare provider immediately if you believe you did not receive a full dose.
Avoid placing fingers near the needle end, as the Needle Depth Adjuster does not lock.
Carefully unscrew the Syringe Housing from the Handle, holding onto the housing with the syringe.
Tip Syringe Housing to drop syringe into Sharps Container; do not use regular trash or recycling bin.
Wipe external parts and accessible areas of Syringe Holder and Handle with a damp cloth.
Do not use immersion, detergents, solvents, or alcohol for cleaning the autoinjector.
Store the Autoinjector in its Protective Case until required for your next dose.
Store in a dry, room-temperature place, out of sunlight; do not refrigerate or store syringe inside.