Environment
Maximum altitude 2,000 m
Pollution Degree 2
Indoor use only.
Operating Environment
Ambient temperature range 0 °C to 55 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 low temperature
limit and MIL-PRF-28800F Class 2 high
temperature limit.)
Relative humidity range 10% to 90%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
Storage Environment
Ambient temperature range -40 °C to 71 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 limits.)
Relative humidity range 5% to 95%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
Caution Clean the PXIe-8398 with a soft nonmetallic brush. Make sure that the
device is completely dry and free from contaminants before returning it to service.
Shock and Vibration
Operating shock 30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC-60068-2-27. Meets MIL-
PRF-28800F Class 2 limits.)
Random vibration
Operating 5 Hz to 500 Hz, 0.3 g
rms
Nonoperating 5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC-60068-2-64. Nonoperating test
profile exceeds the requirements of MIL-
PRF-28800F, Class 3.)
MXI-Express Gen-3 x16 User Manual | © National Instruments | 27

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments MXI-Express Gen-3 x16 and is the answer not in the manual?

National Instruments MXI-Express Gen-3 x16 Specifications

General IconGeneral
BrandNational Instruments
ModelMXI-Express Gen-3 x16
CategoryController
LanguageEnglish

Related product manuals