4. Configure the power sensor with the following settings:
Maximum noise ratio: 0.005 dB
Auto range: Enabled
Power measurement: Continuous average
Path selection: Automatic
Averaging: Automatic
Aperture: 20 ms
Timeout: 10 s
Frequency: Frequency listed in the Passband Flatness Accuracy Verification table
5. Configure the PXIe-5413 with the following settings:
Load impedance: 50 Ω
Output mode: Standard Function
Waveform type: Sine
Amplitude: Amplitude range listed in the Passband Flatness Accuracy Verification
table
Frequency: Frequency listed in the Passband Flatness Accuracy Verification table
Offset: 0 V
6. Wait 1 second for settling.
7. Measure and record the power (dBm) using the power sensor.
When the power is measured at 0.05 MHz, record this as P
Measured at 0.05 MHz
.
When the power is measured at all frequencies other than 0.05 MHz, record this as
P
Measured at Current Test Frequency
.
8. Skip this step if the PXIe-5413 frequency is configured to 0.05 MHz. Calculate the
passband flatness error using the following formula:
Passband Flatness Error (dB) = P
Measured at 0.05 MHz
- P
Measured at Current Test Frequency
where P
Measured at 0.05 MHz
(dBm) and P
Measured at Current Test Frequency
(dBm) are values for
power measured at the same amplitude range listed in the Passband Flatness Accuracy
Verification table.
9. Repeat steps 4 through 8 for each configuration listed in the Passband Flatness Accuracy
Verification table.
10. For each configuration, compare the passband flatness error to the appropriate limit from
the Passband Flatness Accuracy Verification table.
11. If you are verifying a two-channel PXIe-5413, complete the following steps to verify
CH 1:
a) Abort waveform generation on the PXIe-5413.
b) Disconnect the SMA (m)-to-N (f) adapter and power sensor assembly from CH 0
and connect the assembly to CH 1 of the PXIe-5413.
c) Repeat steps 4 through 10.
PXIe-5413 Calibration Procedure | © National Instruments | 13

Other manuals for National Instruments PXIe-5413

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXIe-5413 and is the answer not in the manual?

National Instruments PXIe-5413 Specifications

General IconGeneral
BrandNational Instruments
ModelPXIe-5413
CategoryInverter
LanguageEnglish