EasyManua.ls Logo

National Instruments RMX-4120

National Instruments RMX-4120
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© National Instruments | B-5
RMX Programmable Power Supplies User Manual
Unable to Perform Remote Control
Remote control through the communication interface cannot be performed.
Have you enabled the LAN interface in the CONFIG settings? Set the LAN interface parameter
(CF20) to "on". Also make sure that the Command language parameter (CF21) is set to "SCPi".
When the LAN interface is in use, the IP address cannot be obtained from the DHCP
server.
Is the LAN LED lit in orange or red? If it is lit in orange, the RMX programmable power supply
is waiting for a response from the DHCP server. If it turns red afterward, the RMX
programmable power supply may have timed out. Consult your network administrator.
Unable to Control Functions Properly
The bleeder on/off function does not work.
Are you using master-slave parallel operation or series operation? When using master-slave
parallel operation or series operation, set the bleeder on/off settings (CF11) of all connected units
to the same value.

Table of Contents

Other manuals for National Instruments RMX-4120

Related product manuals