5 Connecting accessories to Aurical Aud
It is a general rule for all electrical equipment used in the proximity of the client that:
The connected equipment must comply with IEC 60601-1 except for the PC, and equipment connected to the line in
and the line out sockets of Aurical Aud.
See also General warnings 45.
For a detailed description of the connection panel, see the Aurical AudReference Manual.
Connection panel - Aurical Aud
A. PC/USB connection
B. Powered USB connections for accessories
C. External power supply
D. Sound field speaker output (optical digital line-out)
E. Sound field speaker output (coaxial digital line-out)
F. Sound field speakers (power output)
G. Insert earphones
H. Headphones - air conduction
I. Patient Responder
J. Bone conductor
K. Speaker, Analog (line output)
L. Line-in
M. Operator monitor headset - headphones
N. Operator monitor headset - boom microphone
O. Counseling and Simulations headphones
P. Talk-back microphone
Note Blue corresponds to Left and red corresponds to Right.
Warning Use only the power supply provided by Natus.
Warning Avoid connecting or disconnecting the hearing instrument while the connector light indicator is lit.
Warning When you connect other electrical equipment to Aurical Aud, remember that equipment that does not
comply with the same safety standards as Aurical Aud can lead to a general reduction in the system's safety level.