What to do if the Natus Medical Equipment ICS AirCal is beeping due to overheating?
- JJennifer KruegerOct 31, 2025
If the ICS AirCal Medical Equipment from Natus is beeping continuously due to overheating, there are several possible causes. First, check the Water Level Display. If the bar is orange, refill the water reservoir following the Filling and Priming procedure. If the bar is blue, there should be sufficient water, but you should still verify. If there is no or very low water refill the water. Another cause might be that the ICS AirCal was tilted 20° or more while the pump was operating, potentially causing the pump to take in air instead of water. In this case, turn off the ICS AirCal, check if there is water present in the reservoir, and follow the Filling and Priming procedure. After addressing these potential causes, turn the unit back on to see if it is working properly. If the issue persi...