Camino
®
ICP Monitor
Intracranial Pressure and Temperature Monitor
User’s Manual
025870 Rev 00 (01/2019)

Table of Contents

Questions and Answers

  • K
    Kathryn TownsendAug 18, 2025
    What to do if natus Camino ICP Medical Equipment shows ICP is out of accuracy range?
    • R
      Ryan RodriguezAug 18, 2025
      If the message 'ICP is out of accuracy range' appears during implantation or removal of the Flex Catheter on your Natus Medical Equipment, disconnect and then reconnect the Flex Catheter to the Flex Extension Cable. If the message appears during patient treatment, do not rely on ICP measurements until the Mean ICP value falls within the accuracy range.
  • M
    Michael DyerAug 20, 2025
    What does 'Temperature is out of accuracy range' mean on natus Camino ICP?
    • A
      amberjohnsonAug 20, 2025
      If your Natus Medical Equipment displays 'Temperature is out of accuracy range', it means the temperature value has fallen under the range that the Camino ICP Monitor can accurately measure (between 30 °C to 42 °C). In this case, do not rely on temperature measurements until the temperature value falls within the accuracy range.
  • R
    Roberto GarciaAug 23, 2025
    What to do if ICP is above alarm limit on my natus Camino ICP Medical Equipment?
    • M
      mikaylarodriguezAug 23, 2025
      If the Mean ICP value from your Natus Medical Equipment has exceeded the user-specified high ICP alarm limit for more than 5 seconds, you should immediately respond to the patient's needs. The alarm will turn off once the patient’s Mean ICP value falls within the alarm limit.
  • J
    Julia SmallAug 26, 2025
    Why is my natus Camino ICP showing 'Monitor overheating, check vent'?
    • S
      Sara EllisAug 26, 2025
      If the temperature inside your Camino ICP Monitor has exceeded 80 °C, it means that your Natus Medical Equipment is overheating. You should check the air vent on the rear panel of the monitor and remove anything that is blocking it.
  • C
    Carla HollowayAug 28, 2025
    What to do if natus Camino ICP says battery is low?
    • Y
      yreynoldsAug 28, 2025
      If your Natus Medical Equipment displays a 'Battery is low' message, it means the battery has 15 minutes or less of charge. In this case, plug the Camino ICP Monitor into an AC power outlet. If the battery has 5 minutes or less of charge, accompanied by an alarm tone, plug the Camino ICP Monitor into an AC power outlet.
  • A
    Alexander JohnsonAug 31, 2025
    What to do if Natus Medical Equipment shows catheter failure?
    • N
      Norman FuentesAug 31, 2025
      If you are getting a 'Catheter failure' message on your Natus Medical Equipment, it could be due to a sudden pressure change during implantation or removal of the Flex Catheter. Disconnect and reconnect the Flex Catheter after completing the implantation or removal process. If the catheter breaks or fails during initialization, disconnect it and replace it with a new one, following the directions for use supplied with the catheter.
  • S
    Sean WatsonSep 2, 2025
    How to fix battery failure on natus Camino ICP Medical Equipment?
    • S
      ssandersSep 2, 2025
      If your Natus Medical Equipment displays 'Battery failure', it could be due to the battery’s control system failing. You can try plugging the Camino ICP Monitor into an AC power outlet. After patient treatment, turn the Camino ICP Monitor off and then back on. If the problem persists, replace the battery, and if the issue still continues, contact Natus Neuro.
  • A
    anthony70Sep 5, 2025
    What to do if Natus Medical Equipment says more than one catheter connected?
    • P
      philliprodriguezSep 5, 2025
      If your Natus Medical Equipment displays 'More than one catheter connected', it means that two ICP catheters types (Fiber Optic and Flex) are connected simultaneously. Remove the ICP catheter that will not be used during patient treatment.
  • P
    Patrick AdamsSep 7, 2025
    What to do if natus Camino ICP shows cooling fan failure?
    • M
      Melinda FuentesSep 7, 2025
      If your Natus Medical Equipment displays a 'Cooling fan failure' message, it means the cooling fan inside the Camino ICP Monitor has stopped running. To prevent overheating, turn off the Camino ICP Monitor and contact Natus Neuro.
  • T
    Tracy ClarkSep 10, 2025
    How to resolve temperature sensor failure on natus Camino ICP Medical Equipment?
    • S
      snavarroSep 10, 2025
      If your Natus Medical Equipment displays 'Temperature sensor failure', it means the temperature catheter is broken. Disconnect the temperature catheter and replace it with a new one, following the directions for use supplied with the catheter.

Summary

Setting Up The System For The First Time

Initial Setup Procedures

Covers unpacking, installing the battery, preparing the power adapter, and initial configuration.

Setting Up The System For Clinical Use

Catheter Connectivity and Operation

Explains catheter types, connection procedures, and operational considerations.

Connecting to a Patient Bedside Monitor

Monitoring Patient ICP and Temperature

Touch Screen Interface and Data Monitoring

Explains the touch screen, viewing ICP/Temp data, scaling, and trend data.

Alarm Management and User Settings

Configuring alarms, customizing settings like waveform, language, time, and date.

Monitoring Trend Data

Trend Data Overview and Extraction

How to view and extract recorded ICP trend data over a 5-day period.

Responding to Alarms

Alarm Types, Symbols, and Prioritization

Explains alarm types, symbols, prioritization, and indicators for various alarm conditions.

Responding to Specific Alarms

Detailed procedures for handling physiological, technical, accuracy, and battery alarms.

Extracting Trend Data

Data Extraction Methods

Overview of options for exporting ICP trend data via USB or digital streaming.

Cleaning and Sterilizing the System

Troubleshooting the System

Testing and Preventative Maintenance

Alarm and Power System Testing

Procedures for testing the high ICP alarm, AC power, and battery charge functionality.

Technical Support and Maintenance

natus Camino ICP Specifications

General IconGeneral
Brandnatus
ModelCamino ICP
CategoryMedical Equipment
LanguageEnglish