Olympic Circumstraint
Instructions
Intended Use
The Olympic Circumstraint is intended for the tempo-
rary immobilization of infants during circumcision
and other procedures, such as transfusions, minor
surgery, x-rays, and placement of scalp IVs.
Using the Circumstraint
Required items:
Circumstraint
Arm straps, 8-in. (x2)
Leg straps, 10-in. (x2)
To install the straps:
1 Thread the arm straps through the upper slots.
2 Thread the leg straps in the lower slots.
3 Ensure that the Velcro® hooks are facing out, away
from the infant’s skin.
Figure 1 Installing the Velcro straps
To immobilize the infant:
1 Place the infant in the tray. Position the straps as
near as possible to the elbows and knees.
2 To fasten the straps: Gently press one end of the
strap on the infant’s arm or leg, while bringing the
opposite end of the strap across the top, then
snugly fasten the Velcro® strap (Figure 2).
Figure 2 Immobilizing the infant
To position for circumcision:
The infant’s perineum should rest against the
raised platform between his legs (Figure 3). This
elevates the genitalia for the procedure.
For smaller infants, place a folded towel under the
hips for additional elevation.
Figure 3 Positioning for circumcision
CAUTION
Only use this device under the direct super-
vision of a licensed medical practitioner.
The infant should be under continuous
observation when immobilized.
Use of non-Olympic straps is not recom-
mended and may lead to product failure.
Damage resulting from using straps
purchased from other manufacturers is not
covered by the product warranty.
Arm straps (x2)
Leg straps (x2)
Question and Answer IconNeed help?

Do you have a question about the natus Olympic Circumstraint and is the answer not in the manual?

Summary

Intended Use

Using the Circumstraint

To install the straps

Instructions for threading and securing the arm and leg straps to the device.

To immobilize the infant

Step-by-step guide to placing and fastening the infant securely using the straps.

To position for circumcision

How to position the infant's perineum against the raised platform for the procedure.

Cleaning the Circumstraint

To clean the straps

Methods for disinfecting and cleaning the device's straps, including laundering.

To clean the board

Instructions for wiping down the main board using disinfectant wipes.

Storing the Circumstraint

Specifications

Ordering Information

Summary

Intended Use

Using the Circumstraint

To install the straps

Instructions for threading and securing the arm and leg straps to the device.

To immobilize the infant

Step-by-step guide to placing and fastening the infant securely using the straps.

To position for circumcision

How to position the infant's perineum against the raised platform for the procedure.

Cleaning the Circumstraint

To clean the straps

Methods for disinfecting and cleaning the device's straps, including laundering.

To clean the board

Instructions for wiping down the main board using disinfectant wipes.

Storing the Circumstraint

Specifications

Ordering Information

Overview

The Olympic Circumstraint is a medical device designed for temporary immobilization of infants during procedures such as circumcision, minor surgery, X-rays, and placement of scalp IVs. It is intended for use by a licensed medical practitioner.

Function Description

The primary function of the Olympic Circumstraint is to safely and temporarily immobilize an infant. This is achieved by securing the infant to a board using a system of adjustable straps. The device helps to maintain the infant in a stable position, facilitating various medical procedures. For circumcision, the device positions the infant with arms extended above the head and legs spread apart, allowing access to the genital area.

Important Technical Specifications

Weight Capacity: The Olympic Circumstraint is designed for infants weighing between 3.9-9.9 lb (1800-4500 g). Product Weight: The device itself weighs less than 2 lb (0.9 kg). Board Dimensions: The board measures 25.25 inches high x 13 inches wide x 3.5 inches deep (64.5 cm high x 33 cm wide x 8.9 cm deep). Materials:

  • Board: Made from dense, non-absorbent plastic, ensuring durability and ease of cleaning.
  • Straps: Made from quick-drying nylon with Velcro® fasteners, providing secure and adjustable immobilization. Latex-Free: Both the Circumstraint board and its straps are not made with natural rubber latex, reducing the risk of allergic reactions.

Usage Features

Immobilization Steps:

  1. Placement: Position the infant on the Circumstraint board, ensuring the straps are as near as possible to the elbows and knees.
  2. Securing Straps: Fasten the straps gently. Begin with one end of the strap on the infant's arm or leg, bringing the opposite end of the strap across the top, then snugly fasten the Velcro® strap. Circumcision Positioning:
  • For circumcision, the infant's perineum should rest against the raised platform between the legs. This elevates the genitalia for the procedure.
  • For smaller infants, a folded towel under the hips can provide additional elevation. Strap System: The device includes various strap sizes to accommodate different infant dimensions:
  • Arm Straps: Two 8-inch (20.3-cm) arm straps are included with the complete unit.
  • Leg Straps: Two 10-inch (25.4-cm) leg straps are included with the complete unit.
  • Replacement Strap Sets:
    • Regular Set (REF 50108): Includes four 8-inch (20.3-cm) arm straps and four 10-inch (25.4-cm) leg straps.
    • Long Set (REF 50109): Includes four 8-inch (20.3-cm) arm straps and four 13-inch (33-cm) leg straps, offering more length for larger infants or specific positioning needs. Safety Precautions:
  • The device is intended for temporary immobilization only.
  • Continuous observation of the infant is required when immobilized.
  • The Olympic Circumstraint is not intended to be used as a restraint device for unattended continuous observation or when immobilized.
  • Use of non-Olympic straps is not recommended as it may lead to product failure. Damage resulting from using straps purchased from other manufacturers is not covered by the product warranty.

Maintenance Features

Cleaning the Circumstraint:

  • Board: Wipe the board using disinfectant wipes.
  • Straps: Disinfect the straps by laundering or soaking them in a liquid disinfectant. Straps can also be sterilized using steam, gas, or pasteurization methods. Cleaning Agents:
  • Only liquid disinfectant should be used.
  • CAUTION: Do not heat sterilize the board. Do not use cleaning agents containing ketones, ether, or plastic solvents as they will damage the board. Storage:
  • The Circumstraint can be hung on a wall using the hole on the top of the device.
  • Multiple Circumstraints can be nested together on a shelf for compact storage.

Ordering Information

For information or ordering, contact Natus Medical at 1-800-303-0306 or +1-650-802-0400.

  • Olympic Circumstraint (Complete unit): REF 50100 (includes two 8-in. arm straps and two 10-in. leg straps).
  • Replacement Strap Set, Regular: REF 50108 (set of eight straps: four 8-in. arm straps and four 10-in. leg straps).
  • Replacement Strap Set, Long: REF 50109 (set of eight straps: four 8-in. arm straps and four 13-in. leg straps).

natus Olympic Circumstraint Specifications

General IconGeneral
Brandnatus
ModelOlympic Circumstraint
CategoryMedical Equipment
LanguageEnglish