9
Special Functions
The functions described on this page must be activated using a
computer and the DOOR conguration software. It is also possible to
order DOOR with these functions acitvated.
Auto Reset Timeout
It is possible to congure DOOR with an auto reset timeout for the
external alarm sensor. When the external sensor is activated, DOOR
waits during the auto reset time before sending an alarm. If the alarm
sensor is activated a shorter time than the auto reset time, no alarm
will be sent.
This function is useful for example with a bed sensor connected to
DOOR. The patient can then go to the toilet without an alarm being
sent as long as he returns to bed within a certain time.
Peek Function
With a carpet sensor connected to DOOR and the peek function
enabled, the door can be opened without any alarm being sent as long
as the patient is standing on the door mat.
Combining the peek function and the auto reset timeout gives the
patient the possibility to go out and fetch the newpaper without an
alarm being sent, as long as he returns to the room and closes the
door from the inside within a certain time.