EasyManua.ls Logo

NEC MD212G3 - Changing the configuration

NEC MD212G3
27 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English-18
3. Switching the input source, press and hold both the buttons for 2 seconds. And the “PRESENT STATE” OSD main menu
with the selected input source appears.
Current signal input source is confi rmed on the INPUT description of “PRESENT STATE” OSD main menu.
Input source
Changing the confi guration (Preset 1, 2, 3)
The NEC model MD212G3 is shipped with 3 factory preset confi gurations, as listed below.
Confi guration
Luminance (cd/m
2
)
Gamma Remark
Maximum Minimum Ambient
1 400 0.6 0.0 DICOM [Factory Default]
2 400 1.0 0.0 DICOM
3 500 0.6 0.0 DICOM
The NEC model MD211G5 is shipped with 3 factory preset confi gurations, as listed below.
Confi guration
Luminance (cd/m
2
)
Gamma Remark
Maximum Minimum Ambient
1 500 1.0 0.0 DICOM [Factory Default]
2 550 1.0 0.0 DICOM
3 400 1.0 0.0 DICOM
The confi guration can be changed via an OSD menu. The confi guration 1 parameters may be modifi ed anytime by a new
display calibration. Regarding display calibration, please refer to the GammaCompMD QA user manual.
HINT: In order to apply a change of confi guration, make sure to press and hold both (+) and (-) buttons for over 2 seconds
after moving the cursor. Otherwise the OSD menu will disappear in 15 seconds and the confi guration will not be
changed.
1. Click the (+) plus button or (-) minus button on the back side of the display once to open the “PRESENT STATE” OSD main
menu. A menu opens on the screen, which looks similar to this picture.
2. With the “PRESENT STATE” OSD main menu displayed, press and hold both the buttons for 2 seconds. The “SWITCHING
CONFIGURATION” submenu appears.
With the screens above displayed, press the power-switch quickly, the OSD submenu switches to “SWITCHING
INPUT SOURCE” and “SWITCHING CONFIGURATION” in that order.