6 Specifications
Product Specifications
Illumination Source L-30 Halogen Bulb (3 V, 2 W)
Correction Range -36 D to +35 D
Filter Lever (Illumination system) Polarizing filter, 4000 K filter
Observation Polarizing Filter ON/OFF
Illumination Dial
Normal Aperture, Small Aperture, Slit, Concentric
Scale, Red-Free filter
Batteries
BXα-13 C-size batteries (2 pcs.)
BXα-13A AA-size batteries (2 pcs.)
Illumination Time Approximately 80 min.
Dimensions
*1
BXα-13
85 mm x 245 mm x 45 mm
Approximately 550 g
BXα-13A
85 mm x 245 mm x 45 mm
Approximately 450 g
*1: Values when stored in case.
Classification
Degree of protection against electric
shock
Internally powered ME equipment
Applied parts
No applied parts
Degree of protection against harmful
ingress of water or particulate matter
IPX0
Method(s) of sterilization
Do not sterilize
Suitability for use in an oxygen rich
environment
Do not use in oxygen rich environments
Mode of operation
Continuous operation
Safety Standards
Electric Safety
IEC 60601-1:2005+AMD1:2012
Electromagnetic Disturbances
IEC 60601-1-2:2014
Ophthalmic Instruments
ISO 15004-1:2006
Light Hazard
ISO 15004-2:2007
Direct Ophthalmoscopes
ISO 10942:2006
Environmental Conditions
Use
Storage
Transport
Temperature
+10 °C to +35 °C
-10 °C to +55 °C
-10 °C to +55 °C
(50 °F to 90 °F)
(14 °F to 131 °F)
(14 °F to 131 °F)
Relative Humidity
30 % to 90 %
10 % to 95 %
10 % to 95 %
(no condensation)
(no condensation)
(no condensation)
Atmospheric
Pressure
800 hPa to 1060 hPa 500 hPa to 1060 hPa 500 hPa to 1060 hPa