EasyManuals Logo
Home>Neslab>Water System>RTE-Series

Neslab RTE-Series User Manual

Neslab RTE-Series
27 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
- 20 -
Reservoir Cleaning
Routine cleaning can be achieved by simply sponging down the seamless
stainless steel tank with tap water. (Dish washing detergent may be used
but the tank must be thoroughly rinsed.)
To gain access to the entire reservoir the pump box and reservoir cover
should be removed.
Remove the line cord from the rear of the unit and then remove the four
screws (two on each side) securing the reservoir's cover.
To get slack on the refrigeration cable, remove the cable strap. The cable
itself does not need to be disconnected.
Turn the cover assembly over and carefully place it on a supporting plat-
form.
Algae
To restrict the growth of algae in the bath, we recommend the bath cover be
kept in place and that all circulation lines be opaque. This will eliminate the
entrance of light required for the growth of most common algae.
NESLAB recommends the use of Chloramine-T, 1 gram per 3½ liters.
Refrigeration Cable
Cable Strap

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Neslab RTE-Series and is the answer not in the manual?

Neslab RTE-Series Specifications

General IconGeneral
BrandNeslab
ModelRTE-Series
CategoryWater System
LanguageEnglish

Related product manuals