NESS L300 User Guide 45
Water protection
IPX1
Environmental
:
Transport & Storage: 0 – 35 Cº
Operating: 0-45 Cº
Charging: 10-45 Cº ( 50-113 F º )
10% to 85% relative humidity
900hPa to 1060hPa atmospheric pressure
Orthosis Electronic Module Specifications
Classification
Internally powered, continuous operation with type BF applied
parts
Operating voltage
:
3.7 V
Operation current
:
70 mA
Battery type
:
Rechargeable Prismatic LiIon (Lithium Ion) 3.7V, 700 mAh
Charging time
:
3 Hours (typical) using provided power supply
Charging cycles
:
At least 350 charging cycles
Operation period
:
At least 2 days between charges
Battery discharge
time when stored
:
At least 1 week between charges
Indications
:
Status (fault, battery, charging) & Stimulation LEDs
Buzzer for audible indications
Pulse Parameters
Drive method: Constant current
Intensity 0-80 mA, 1 mA resolution (measured at positive phase)
Shape: Symmetric Asymmetric