Ensuring cable connection
If one of the cables is disconnected from the Trial Device, it will beep continuously. Reinsert the cable using the following steps:
1. Ensure that stimulation is OFF. Press to stop the alarm and stimulation.
2. Check the color of the sticker at the
Trial Device end of the OR cable.
3. Plug the OR cable into Trial Device port that matches the color of the OR cable sticker. A click will sound once the cable is
fully inserted into the Trial Device.
4. Use Remote Control to turn stimulation back on.
Replacing the batteries
WhentheLowBatteryIndicatorLightisashingorON,theTrialDevice’s batteries need to be replaced.
To change the batteries, follow these steps:
1. Turn the Trial Device OFF by pressing .
2. Remove the battery cover on the back of unit and remove the 2 batteries.
3. Conrmthenew batteries are the purple batteries provided by Nevro
®
. (These batteries should not be used for any other
purpose.) DO NOT use AA batteries.
4. Insert the new purple batteries into the battery compartment. Be sure to match (+) and (–) ends of batteries with the
corresponding connectors inside battery compartment.
5. Replace the cover and restart stimulation using the Remote Control.
Turning stimulation ON
• Refer to the Remote Control Guide.
Changing stimulation level
• Refer to the Remote Control Guide
.