Tempest Nd:YAG Laser Operator’s Manual
20
The power supply has been preset at the factory for 95 – 125 VAC or
200 – 250 VAC single phase operation, depending upon the voltage
supply available at your location. However, the laser can be switched on
site from 95-125 VAC to 200-250 VAC. In addition, the fuses must be
changed to match the voltage setting.
Figure 3-1: Voltage Switch
WARNING: Make sure that the switch is flipped to the correct
voltage. In addition, both fuses must be changed to match the
current voltage setting.
The AC power entry module located on the back of the power supply
holds two power line fuses. For 95 – 125 VAC operation two 10
amp/250V, 5 x 20 mm, time lag, high breaking capacity fuses are
installed. For 200 – 250 VAC operation two 6.3 amp/250V SPT 5 x 20
mm time lag, high breaking capacity fuses are installed. Acceptable
fuses are:
Manufacturer
90–125 VAC 200–250 VAC
Schurter 001.2514 001.2512
The high voltage switching power supply, located on the top shelf of the
power supply, has two PC board mounted type 3AG fuses. One fuse,
labeled F1, is a 10A 32 V SloBlo, while F2 is a 1/32 A 250 V SloBlo
fuse. Acceptable fuses are:
Manufacturer
Fuse 90–250 VAC
Littelfuse F1 313.010
Littelfuse F2 313.031
Voltage
Requirements
Power Line
Fuse
Voltage Switch