Troubleshooting
- Page 59 -
Adjust the 4HG crystal angle to maximize the output energy.
Adjust the wave plate on the input to the SHG housing to maximize
the energy.
Adjust the SHG crystal angle to maximize the output energy.
If the optional attenuator is installed, ensure that the attenuator knob
is rotated fully clockwise to the maximum energy/minimum
attenuation setting.
Check the harmonic crystals and housings to ensure that all surfaces
are clean and undamaged.
If none of the above work, check the 532 nm energy, then the 1064
nm energy.