SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
D. Remove Air Intake Tube Connector
See Figure 4.4, Move the Air Intake Tube out in direction of arrow 2 after disconnecting Air
Intake Tube from Air Tube Connector in direction of arrow 1 .
4-3
I2
FIGURE 4.4
E. See figure 4.6, take out the Humidity baffle. See step P in section 2.3, remove 2pcs bolts
on the air filter cover, and take out the air filter.
F. Remove the Access door gasket and Access door cuff, and Iris port sleeve from hood.
G. Remove the soft tubing from hood.
4.2.2 CLEANING PROCEDURE
CAUTION: Some chemical cleaning agents may be conductive and/or leave a residue
which may permit a build-up of dust or dirt which may be conductive. Do not
permit cleaning agents to contact electrical components. Do not spray
cleaning solutions onto any of those surfaces.
A. Clean the skin temperature sensor
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or
paper towel.
CAUTION: Do not steep Sensor into disinfectant-detergent.
B. Clean the Soft tubing, Main Deck Seal, Gaskets, Iris Port Sleeves, Air Intake Tube
Allow parts to soak as recommended by the cleaning solution’s manufacturer, then remove
them and dry completely with a clean cloth or paper tower.
C. Clean the Hood
Use a disinfectant-detergent to thoroughly clean all surfaces, including all holes, indentations,
Access Doors and Iris Ports then dry them with a clean cloth or paper towel.
CAUTION: Alcohol can cause crazing of the clear Acrylic panel. Do not use alcohol,
acetone, or any organic solvents for cleaning. Do not expose the panel
assembly to direct ultraviolet radiation.