Chapter 3 Nasal NO measurements
8 002346-02 NIOX VERO
®
Nasal User Manual English
Note: Performing a nasal measurement without using the NIOX Panel
(see section 4.3, Perform Nasal measurements with NIOX® Panel)
will not produce a graphical result, it will only give the numerical
result of nasal nitric oxide measurement on the NIOX VERO screen.
CAUTION! Make sure to use the correct size olives, the seal between the
olive and nostril must be airtight to ensure correct measurement result.
(A) Nasal Olive Adult, (B) Nasal Olive Pediatric
CAUTION! Do not insert the Nasal Olive inside the nostril. The larger
part of the Nasal Olive should remain outside the nostril.
CAUTION! Ensure that the sampling hole in the Nasal Olive is aligned
with the nasal passage and not blocked. A blocked sampling hole will in-
fluence the measurement result.
Note: If an error occurs see “Alert codes and actions” on page 11.