Chapter 5 Cleaning procedure
19
5 Cleaning procedure
WARNING! DO NOT clean the instrument or handle with
products containing alcohol. This includes sprays or wipes
containing alcohol!
WARNING! DO NOT clean area immediately surrounding the
NIOX VERO
®
with products containing alcohol. This includes
sprays or wipes containing alcohol.
For cleaning of the instrument see NIOX VERO User Manual.
Note: NIOX Nasal Kit are disposable items for single patient use only.
• If the any part of the Nasal Kit appears clogged, replace the Nasal Kit
• The NIOX Nasal Kit is for single patient use and do not need any
cleaning.
6 NIOX VERO
®
Nasal
Technical Specification
6.0.1 Electromagnetic emissions
CAUTION! This equipment has been designed and tested to CISPR 11
Class A. In a domestic environment it may cause radio interference, in which
case, you may need to take measures to mitigate the interference.
For guidance and manufacturer’s declaration - electromagnetic emissions
see www.niox.com
6.1 Performance data
Measurement range: 5 to 2000 ppb
Lowest detection limit: 5 ppb
Sampling flow rates: 5 ml/s
6.2 Linearity
Squared correlation coefficient r2 ≥ 0.998
6.3 Precision
< 3ppb of measured value for values < 30 ppb, < 10% of measured value for
values ≥ 30 ppb. Expressed as one standard deviation for replicate mea-
surements with the same instrument, using a certified gas concentration of
Nitric Oxide reference standard.
6.4 Accuracy
±10 ppb for measured values < 50ppb or 20% of measured value for values
≥ 50 ppb. Expressed as the upper 95% confidence limit, based on absolute
mean of differences from certified gas concentration of Nitric Oxide.