Electromagnetic Immunity Declaration
NIRA Pro is intended for use in the electromagnetic environment specified below. The customer or the user of NIRA Pro should assure that it is used in such an environment.
Immunity test IEC 60601test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
Complies
Electrical fast transient/ burst
IEC 61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines
Complies
Mains power quality should be that of a typical home, com-
mercial or hospital environment.
Surge
IEC 61000-4-5
+1 kV differential mode
+2 kV common mode
Complies
Mains power quality should be that of a typical home com-
mercial or hospital environment.
Voltage dips, short interrupts and
voltage variations on power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
For 0,5 cycle 40 % UT
(60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
Complies
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m Complies
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical home, com-
mercial or
hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.