EasyManua.ls Logo

NIX QC - Specifications

NIX QC
13 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20 21
NIX QC: MANUAL
Specifications
Size
Weight
Measurement interface
Plastic enclosure
Other physical attributes
Carrying case attributes
Inter-instrument agreement
Short-term repeatability
Calibration/validation
Spectral range
Measurement geometry
Measurement aperture
Light source
Temperature range
Humidity range
Time per scan
Bluetooth
Charging interface
Charge time
Battery capacity
Light source estimated
lifespan
6.0 x 6.0 x 4.3 cm
50 g
Blocks all ambient light
2 mm thick polycarbonate ABS blend
Heat staked; ultrasonically welded; no moving
parts
Water resistant; portable; durable; eld-tested
Average ΔE (CIE2000) < 0.30, maximum ΔE
(CIE2000) < 0.75 (both D50, 2° & D65, 10°)
to master unit on 24 reference tiles at a
temperature of 23 °C
ΔE (CIE2000) < 0.10 (both D50, 2° & D65, 10°)
to mean of 10 scans taken every 5 seconds
on included reference tile at a temperature
of 23 °C
On an array of proprietary 24 colored tiles
at Nix Sensor Ltd.s facility (comprehensive,
extended gamut)
380-730 nm
45°/0°
14 mm diameter
2x 5000K and 2x 6500K High-CRI LEDs
designed specically for color reproduction
5 °C 35 °C
0% 80% non-condensing
< 2.5 seconds
Bluetooth Low Energy 4.0, range 20 m line of
sight
Micro USB 2.0
4 hours
25,000 scans before recharge; 500 mAh
1,000,000+ scans