7.2. File names
LOG-files:
• The measuring data are saved as „XXXXLOG.txt“ file, whereas XXXX
correspond to the sample number.
Protocol-files :
Stable and actual protocols are saved as „XXXXPRO.txt“ file. The stable
protocol is generated automatically as soon as the stability is reached. If
required the user can generate additional (actual) protocols. For that purpose
the [Start/Stop]-key has to be pressed longer time, whereby the actual protocol
is generated.
The last lines of the protocol contain an electronic signature by what the PC
program NovaLog MC is able to prove the authenticity of the protocol. A
maximum of 255 protocols can be output in the same protocol file!
Data format :
Detailed information concerning the data format are available on demand at
Novasina AG.
8. The battery
To the LabSwift-aw instrument there is a Lithium-Ion battery available. This
allows the operation of the instrument at places without power supply for over
20 hours.
8.1. Operation
If the battery is low the information „LOW BATT“ is displayed and the
instrument gives 5 short beep. In this state there is normally enough time to
finish the current measurement. Certainly the charging can be directly started
by connecting the line adapter.
Is the battery totally discharged, the message „LOW BATT“ will be displayed
for 2 seconds followed by a permanent beep and the instrument is switched off.
Note:
Do not let the instrument in this status (with discharged battery) standing
around for longer time. The battery might be damaged.
8.2. Charging
The charging can be done while the instrument is switched on or off. Ideally the
charging is made when the instrurment is switched off to prevent a heating of
the chamber temperature. The charge of battery is indicated with a 2-colour
LED on the backside of the LabSwift-aw :
• red the battery is charging
• green the battery is fully charged