4. Description of Instrument Functions
The LabSwift-aw system is a simple to use laboratory measurement instrument,
which can be powered with a Lithium-Ion rechargeable battery (option). The
device can be adjusted to the user’s needs by the parameter settings explained
in this chapter.
4.1. Measurement Instrument / Operation
4.1.1. Starting the system
The system performs an extensive function check of the internal modules as
well as of important software function during the start-up and operation.
Observed failures are shown on the display.
After a warm-up phase the device is ready and the display switches over to the
measurement mode.
In the measurement mode the following functions are available :
- Button [Start/Stop]: Start the measurement
o Starts a measurement -> „ANALYZE“
It is recommended to start the „ANALYZE“
function after the insert of each sample.
Note:
The instrument can be configured, in order to start „ANALYZE
automatically after closing the cover. See also chapter 4.2.2.
Simultaneously with the beginning of the measurement also the LOG function
on the SD card get started.
o Stops ongoing measurements without stopping the LOG- function.
Note:
An ongoing measurement is stopped together with the LOG function as soon
as the measurement chamber is opened.
Button [Start/Stop] (push button for a longer time):
Storage of the actual protocol
o A protocol with the current measurement values is stored on
the SD-card.
o Per each measurement, which means that max. 255 protocols
can be stored.